Whole-cell tools

Ephys solution calculator

EPSC Internal (low chloride)

Cs⁺/methanesulfonate, low-chloride formulation for isolating excitatory synaptic currents. Evidence & sources →

Powder · weigh Stock · pipette
CompoundFinal mMMW g/molSourceStock mMAddRemove
—1539.0 mg
5000 µL
500 µL
50 µL
—105.9 mg
—8.5 mg
—102.0 mg

Low-chloride EPSC internal solution

This cesium-based internal is intended for voltage-clamp recordings of excitatory postsynaptic currents. Intracellular chloride influences the chloride reversal potential. Interpret current polarity using the holding potential, external solution, and internal composition; the internal recipe alone does not isolate a receptor-mediated current.

Hamill et al. (1981): whole-cell patch-clamp methods

Why ATP and GTP are included

ATP supports energy-dependent cellular processes, while GTP supports G-protein signaling. Check the exact nucleotide salt, hydration state, and assay on the product label before weighing. The molecular weight must match the reagent used.

When should I measure pH?

Check final pH after every ingredient, including ATP and GTP, has dissolved. Equilibrate to the protocol’s specified measurement temperature and record it. Nucleotide additions can change pH; the size of the change depends on the salt forms and buffering.

HEPES-buffered solutions become more alkaline when cooled. A HEPES-only estimate using −0.014 pH units per °C gives an increase of about 0.29 when cooling from 22 °C to 1 °C. This is an approximation, not a correction to apply automatically to a complete internal solution.

HEPES temperature coefficient: manufacturer technical sheet

Do ATP and GTP require preparation on ice?

Cold storage guidance does not establish a universal requirement to perform every preparation step on ice. Use the handling conditions validated for your nucleotide products and complete internal formulation. The stability of a complete solution during an hour at room temperature should not be assumed from powder shipping conditions. Keep final pH measurement temperature separate from storage and aliquoting temperature.

Mg-ATP handling and solution stability: manufacturer technical sheet