Whole-cell tools

Ephys solution calculator

IPSC Internal (high chloride)

CsCl-based, high-chloride formulation for sensitive inhibitory-current detection. Evidence & sources →

Powder · weigh Stock · pipette
CompoundFinal mMMW g/molSourceStock mMAddRemove
—1178.8 mg
5000 µL
500 µL
50 µL
—105.9 mg
—8.5 mg
—102.0 mg

High-chloride IPSC internal solution

This high-chloride internal changes the chloride driving force for inhibitory postsynaptic current recordings. The expected current direction depends on the holding potential and external chloride as well as the internal solution. Keep these conditions in the recording record when comparing IPSCs.

Coombs et al. (1955): ionic mechanisms of inhibition

Why ATP and GTP are included

ATP supports energy-dependent cellular processes, while GTP supports G-protein signaling. Check the exact nucleotide salt, hydration state, and assay on the product label before weighing. The molecular weight must match the reagent used.

When should I measure pH?

Check final pH after every ingredient, including ATP and GTP, has dissolved. Equilibrate to the protocol’s specified measurement temperature and record it. Nucleotide additions can change pH; the size of the change depends on the salt forms and buffering.

HEPES-buffered solutions become more alkaline when cooled. A HEPES-only estimate using −0.014 pH units per °C gives an increase of about 0.29 when cooling from 22 °C to 1 °C. This is an approximation, not a correction to apply automatically to a complete internal solution.

HEPES temperature coefficient: manufacturer technical sheet

Do ATP and GTP require preparation on ice?

Cold storage guidance does not establish a universal requirement to perform every preparation step on ice. Use the handling conditions validated for your nucleotide products and complete internal formulation. The stability of a complete solution during an hour at room temperature should not be assumed from powder shipping conditions. Keep final pH measurement temperature separate from storage and aliquoting temperature.

Mg-ATP handling and solution stability: manufacturer technical sheet